Digital RS485 Output COD BOD TOC TURBIDITY Sensor

Short Description:

COD sensor is a UV absorption COD sensor, combined with a lot of application experience, based on the original basis of a number of upgrades, not only the size is smaller, but also the original separate cleaning brush to do one, so that the installation is more convenient, with higher reliability.

It does not need reagent, no pollution, more economic and environmental protection.On-line uninterrupted water quality monitoring.Automatic compensation for turbidity interference, with automatic cleaning device, even if long-term monitoring still has excellent stability.


Product Detail

Product Tags

Testing principle:

 

Many organic compounds dissolved in water are absorbent  to  ultraviolet  light. Therefore, the total amount of organic pollutants in the water can be measured by measuring the extent to which these organics absorb ultraviolet light at  254nm.

 

Sensor features:

 

Digital sensor, RS-485 output, support Modbus

No reagent, no pollution, more economic and environmental protection Automatic compensation of turbidity interference, with excellent test performance

With self-cleaning brush, can prevent biological attachment, maintenance cycle more

 

Technical parameters

 

Name Parameter
Interface Support RS-485, MODBUS protocols
COD Range 0.1 to 1500mg/L equiv.KHP
BOD Range 0.1 to 900mg/L equiv.KHP
COD/BOD Accuracy <5% equiv.KHP
COD/BOD Resolution 0.01mg/L equiv.KHP
TOC Range 0.1 to 750mg/L equiv.KHP
TOC Accuracy <5% equiv.KHP
TOC Resolution 0.1mg/L equiv.KHP
Tur Range 0.1-4000 NTU
Tur Accuracy <3% or 0.2NTU
Tur Resolution 0.1NTU
Temperature Range +5 ~ 45℃
Housing IP Rating IP68
Maximum pressure 1 bar
User Calibration one or two points
Power Requirements DC 12V +/-5%,current<50mA(without wiper)
Sensor OD 32 mm
Sensor Length 200 mm
Cable Length 10m (default)