Blue-Green Algae Online Analyzer T6401
Industrial Blue-Green Algae Online Analyzer is an online water quality monitor and control instrument with microprocessor. It is widely used in power plants, petrochemical industry, metallurgical electronics, mining, paper industry, food and beverage industry, environmental protection water treatment, aquaculture and other industries. The Blue-Green Algae value and temperature value of water solution are continuously monitored and controlled.
Blue-green algae online monitoring of water plant inlet,drinking water source,aquaculture and etc.
Blue-green algae online monitoring of different water bodies such as surface water, scenic water, etc.
85~265VAC±10%,50±1Hz, power ≤3W;
9~36VDC, power consumption≤3W;
Blue-green algae: 200—300,000cells/ML
Blue-Green Algae Online Analyzer T6401
Measurement mode
Calibration mode
Trend chart
Setting mode
1.Large display, standard 485 communication, with online and offline alarm, 144*144*118mm meter size, 138*138mm hole size, 4.3 inch large screen display.
2.The data curve recording function is installed, the machine replaces the manual meter reading, and the query range is arbitrarily specified, so that the data is no longer lost.
3.Carefully select materials and strictly select each circuit component, which greatly improves the stability of the circuit during long-term operation.
4.The new choke inductance of the power board can effectively reduce the influence of electromagnetic interference, and the data is more stable.
5.The design of the whole machine is waterproof and dustproof, and the back cover of the connection terminal is added to extend the service life in harsh environments.
6.Panel/wall/pipe installation, three options are available to meet various industrial site installation requirements.
Electrical connection The connection between the instrument and the sensor: the power supply, output signal, relay alarm contact and the connection between the sensor and the instrument are all inside the instrument. The length of the lead wire for the fixed electrode is usually 5-10 meters, and the corresponding label or color on the sensor Insert the wire into the corresponding terminal inside the instrument and tighten it.
Measurement range | 200—300,000cells/ML |
Measurement unit | cells/ML |
Resolution | 25cells/ML |
Basic error | ±3% |
Temperature | -10~150℃ |
Temperature Resolution | 0.1℃ |
Temperature Basic error | ±0.3℃ |
Current Output | 4~20mA,20~4mA,(load resistance<750Ω) |
Communication output | RS485 MODBUS RTU |
Relay control contacts | 5A 240VAC,5A 28VDC or 120VAC |
Power supply (optional) | 85~265VAC,9~36VDC,power consumption≤3W |
Working conditions | No strong magnetic field interference around except the geomagnetic field. |
Working temperature | -10~60℃ |
Relative humidity | ≤90% |
IP rate | IP65 |
Instrument Weight | 0.8kg |
Instrument Dimensions | 144×144×118mm |
Mounting hole dimensions | 138*138mm |
Installation methods | Panel,Wall mounted,pipeline |
Chlorophyll Sensor
Based on the Fluorescent measuring target parameter of pigment,can be identified before affected by potential water bloom.
Without extraction or other treatment, rapid detection to avoid the impact of long shelving the water sample.
Digital sensor,high anti-jamming capacity and far transmission distance.
Standard digital signal output,can achieve integration and networking with other equipment without controller.
Plug-and-play sensors,quick and easy installation.
Measurement range | 200—300,000cells/ML |
Measurement Accuracy | ±10% of the signal level corresponding value of 1ppb Rhodamine B Dye |
Repeatability | ±3% |
Resolution | 25cells/ML |
Pressure range | ≤0.4Mpa |
Calibration | Deviation value calibration,slope Calibration |
Requirements |
Suggest a multipoint monitoring for the distribution of Blue-Green Algaein water is very uneven.Water turbidity
is below 50NTU. |
Main material |
Body:SUS316L(fresh water),Titanium alloy(marine);
Cover:POM;Cable:PUR |
Power supply | DC:9~36VDC |
Storage temperature | -15-50℃ |
Communication protocol | MODBUS RS485 |
Measuring temperature | 0- 45℃(Non-freezing) |
Dimension | Dia38mm*L 245.5mm |
Weight | 0.8KG |
Protective rate | IP68/NEMA6P |
Cable length | Standard:10m,the maximum may be extended to 100m |