Product Principle:
After mixing water sample and masking agent,total nitrogen in the form of free ammonia or ammonium ion in alkaline environment and in the presence of sensitizing agent reacts with potassium persulfate reagent to form a colored complex.The analyzer detects the colour change and converts the change into ammonia nitrogen value and output it.The amount of coloured complex formed is equal to the amount of ammonia nitrogen.This method is suitable for wastewater with total nitrogen in the range of 0-50mg/L. Excessive calcium and magnesium ions, residual chlorine or turbidity may interfere with the measurement.
Technical Parameters:
No. |
Name |
Technical Parameters |
1 |
Range |
Suitable for wastewater with total nitrogen in the range of 0-50mg/L. |
2 |
Test Methods |
Spectrophotometric determination of potassium persulfate digestion |
3 |
Measuring range |
0~50mg/L |
4 |
Detection Lower limit |
0.02 |
5 |
Resolution |
0.01 |
6 |
Accuracy |
±10% or ±0.2mg/L(take the larger value)) |
7 |
Repeatability |
5% or 0.2mg/L |
8 |
Zero Drift |
±3mg/L |
9 |
Span Drift |
±10% |
10 |
Measurement cycle |
The minimum test cycle is 20 minutes. Color chromogenic time can be modified in 5-120min according to site environment. |
11 |
Sampling period |
Time interval (adjustable), integral hour or trigger measurement mode can be set. |
12 |
Calibration cycle |
Automatic calibration (1-99 days adjustable), according to actual water samples, manual calibration can be set. |
13 |
Maintenance cycle |
Maintenance interval is more than one month, about 30 minutes each time. |
14 |
Human-machine operation |
Touch screen display and instruction input. |
15 |
Self checking protection |
Working status is self-diagnostic, abnormal or power failure will not lose data. Automatically eliminates residual reactants and resumes work after abnormal reset or power failure. |
16 |
Data storage |
No less than half a year data storage |
17 |
Input interface |
Switch quantity |
18 |
Output interface |
Two RS232 digital output, One 4-20mA analog output |
19 |
Working Conditions |
Working indoors; temperature 5-28℃; relative humidity≤90% (no condensation,no dew) |
20 |
Power Supply and Consumption |
AC230±10%V, 50~60Hz, 5A
|
21
|
Dimensions | 355× 400×600(mm) |